Введение

Found on page 85 in the full service manual.

  1. Remove the ribbon cable at the connection on the oxygen blender control board.
    • Remove the ribbon cable at the connection on the oxygen blender control board.

  2. Mark the plastic hoses attached to the oxygen blender (so you can put them back on the same way) and then remove.
    • Mark the plastic hoses attached to the oxygen blender (so you can put them back on the same way) and then remove.

  3. The air/oxygen blender has two screws close to the control board. Remove them and remove the unit.
    • The air/oxygen blender has two screws close to the control board. Remove them and remove the unit.

  4. Remove the pneumatic and electrical connections labeled.
    • Remove the pneumatic and electrical connections labeled.

    • Remove the water trap bowls on the rear of the unit.

  5. Remove the four screws and remove the manifold.
    • Remove the four screws and remove the manifold.

    • Remove the two screws holding the power supply to the manifold.

Заключение

To reassemble your device, follow these instructions in reverse order. Calibrate and test the system to see if the replacement was successful

William Burns

Участник с: 06.04.2020

285 Репутация

Автор 2 руководств

Команда

Cal Poly, Team S7-G12, Paton Fall 2020 Участник Cal Poly, Team S7-G12, Paton Fall 2020

CPSU-PATON-F20S7G12

2 членов

Автор 6 руководств

Комментариев: 0

Добавить комментарий

Просмотр статистики:

За последние 24часов: 0

За последние 7 дней: 1

За последние 30 дней: 2

За всё время: 52