For more information on replacing the expiratory valve, please refer to section 10 of the service manual.

  1. Remove 4 Hex Allen screws (A)  from the Rail surrounding the Ventilation Unit.
    • Remove 4 Hex Allen screws (A) from the Rail surrounding the Ventilation Unit.

  2. Remove the Rail (B).
    • Remove the Rail (B).

  3. The Standoff Brackets (C) remain with the Main Cover.
    • The Standoff Brackets (C) remain with the Main Cover.

    • Place aside in a safe location.

  4. Lift the Main Cover (A) from the Ventilation Unit using the Standoff Brackets (B) attached to the Main Cover.
    • Lift the Main Cover (A) from the Ventilation Unit using the Standoff Brackets (B) attached to the Main Cover.

    • Place aside in a safe location.

  5. Disconnect 2 cables from the magnet of the Expiratory Valve.
    • Disconnect 2 cables from the magnet of the Expiratory Valve.

  6. Remove the Expiratory Valve Cover (A) and Silicon Membrane (A) (if not already removed).
    • Remove the Expiratory Valve Cover (A) and Silicon Membrane (A) (if not already removed).

  7. Remove 3 Hex Allen screws (B) M3x8
    • Remove 3 Hex Allen screws (B) M3x8

  8. Remove the Expiratory Valve.
    • Remove the Expiratory Valve.

Заключение

To reassemble your device, follow these instructions in reverse order.

Sierra Saenz

Участник с: 09.04.2020

351 Репутация

Автор 3 руководств

Команда

Cal Poly, Team S7-G25, Paton Spring 2020 Участник Cal Poly, Team S7-G25, Paton Spring 2020

CPSU-PATON-S20S7G25

6 членов

Автор 11 руководств

Комментариев: 0

Добавить комментарий

Просмотр статистики:

За последние 24часов: 0

За последние 7 дней: 1

За последние 30 дней: 2

За всё время: 94