For more information on replacing the motherboard, please refer to section 10 in the service manual.

  1. Remove 4 Hex Allen screws (A)  from the Rail surrounding the Ventilation Unit.
    • Remove 4 Hex Allen screws (A) from the Rail surrounding the Ventilation Unit.

  2. Remove the Rail (B).
    • Remove the Rail (B).

  3. The Standoff Brackets (C) remain with the Main Cover.
    • The Standoff Brackets (C) remain with the Main Cover.

    • Place aside in a safe location.

  4. Lift the Main Cover (A) from the Ventilation Unit using the Standoff Brackets (B) attached to the Main Cover.
    • Lift the Main Cover (A) from the Ventilation Unit using the Standoff Brackets (B) attached to the Main Cover.

    • Place aside in a safe location.

  5. First, we need to remove the ESM module. It is connected to 2 connector sockets. Disconnect and remove.
    • First, we need to remove the ESM module. It is connected to 2 connector sockets. Disconnect and remove.

    • Take out the Ventilation Real-time Controller (VRC) SW chip

  6. Install the ESM Module, by connecting the connectors (A),  and the VRC SW chip on the new board.
    • Install the ESM Module, by connecting the connectors (A), and the VRC SW chip on the new board.

  7. Disconnect all cables and connections from the board to the unit.
    • Disconnect all cables and connections from the board to the unit.

  8. Remove the three Hex screws (B).
    • Remove the three Hex screws (B).

  9. Move the motherboard out then up away from the ventilation unit (C).
    • Move the motherboard out then up away from the ventilation unit (C).

Заключение

To reassemble your device, follow these instructions in reverse order.

Dane Dormedy

Участник с: 08.04.2020

349 Репутация

Автор 3 руководств

Команда

Cal Poly, Team S7-G25, Paton Spring 2020 Участник Cal Poly, Team S7-G25, Paton Spring 2020

CPSU-PATON-S20S7G25

6 членов

Автор 11 руководств

Комментариев: 0

Добавить комментарий

Просмотр статистики:

За последние 24часов: 0

За последние 7 дней: 0

За последние 30 дней: 6

За всё время: 64