Philips BiPAP S/T Air Filter Replacement

Медицинское оборудование
Ни iFixit, ни данное руководство по ремонту не одобрены и не связаны с производителем медицинского оборудования.

Видео обзор

  1. Philips BiPAP S/T Air Filter Replacement, Air Filter: шаг 1, изображение 1 из 1
    • Disconnect the device from the power source to stop airflow.

    • To avoid electrical shock, always unplugged the power cord from the wall outlet before cleaning the device.

  2. Philips BiPAP S/T Air Filter Replacement: шаг 2, изображение 1 из 2 Philips BiPAP S/T Air Filter Replacement: шаг 2, изображение 2 из 2
    • Remove the filter from the enclosure by gently squeezing the filter in the center and pulling it away from the device.

    • Examine the filter for cleanliness and quality.

    • If the foam filter is torn, replace it. (Only Philips Respironics-supplied filters should be used as replacement filters.)

    • A dirty inlet filter may cause high operating temperatures that may affect device performance. Regularly examine the inlet filter.

  3. Philips BiPAP S/T Air Filter Replacement: шаг 4, изображение 1 из 1
    • Wash the gray foam filter in warm water with a mild detergent.

    • Rinse thoroughly to remove all detergent residue.

    • Allow the filter to air dry completely before reinstalling it.

    • Never install a wet filter into the device. You must ensure sufficient drying time.

  4. Philips BiPAP S/T Air Filter Replacement: шаг 5, изображение 1 из 1
    • Reinstall the filter.

    • To install, insert gray foam filter into the filter area.

    • A properly installed, undamaged filter is required for proper operation.

Заключение

To reassemble your device, follow these instructions in reverse order.

Sophia James

Участник с: 04/07/20

193 Репутация

1 руководство автора

Команда

Cal Poly, Team S7-G3, Paton Spring 2020 Участник Cal Poly, Team S7-G3, Paton Spring 2020

CPSU-PATON-S20S7G3

3 членов

Автор 3 руководств

2 Комментариев

In India corona virus control in a earlier situation and now its spread in many countries again '''bipap machines''' are necessary for many peoples in this situation but you have to be careful from your side as well btw thanks for informative article.

mers - Ответить

Why is there a gray blue material coming out on the output filter. It has been doing for over two years. Is it killing me. My bi pa was recalled over a year ago, when are they going to replace it ?

EDward Stone - Ответить

Добавить комментарий

Просмотр статистики:

За последние 24часов: 2

За последние 7 дней: 7

За последние 30 дней: 21

За всё время: 1,299