Перейти к основному содержанию
Присоединяйся к сообществу ремонтников - Создай аккаунт

The ResMed VPAP Auto 25, also known as the ResMed S8 VPAP Auto 25, is a bilevel positive airway pressure (BiPAP) device used to treat sleep apnea. This page provides repair and maintenance information, including common issues and component-level guides.

Receiving system error 1018

Receiving System error 1018 to call service. Unplugged and waited 5-10 minutes, machine still doesn’t work.

Thank you,

Camille Kutis

Ответ на этот вопрос У меня та же проблема

Это хороший вопрос?

Счет 0
Добавить комментарий

1 ответ

Наиболее полезный ответ

@corkie The following message is displayed on the LCD: "SYSTEM ERR 1018 Call service!"

The error is a Flow sensor fault. The solution is to Replace flow sensor. You can download the service manual from this page ResMed VPAP Auto 25 to work on your CPAP. Ultimately you do want somebody (preferably a BioMed person) look it over after you do your repair.

Был ли этот ответ полезен?

Счет 1
Добавить комментарий

Добавьте свой ответ

Camille Kutis буду вечно благодарен.
Статистика просмотров:

За последние 24 часов: 0

За последние 7 дней: 5

За последние 30 дней: 21

За всё время: 743